Tips

Flow Cytometry Tips

We are dedicated to helping our customers achieve exceptional results. While the information below can never be an exclusive solution to every problem you may encounter; it is our hope that you will find the information useful and even beneficial in troubleshooting any problems you may have. Of course, if you find you are still having problems you may submit them to us using this Flow Cytometry Troubleshooting Form.

SYMPTOMS


No signal/weak fluorescence

Signal not correctly compensated

Make sure everything is gated properly by checking the positive single color control and adjust compensation accordingly.

Insufficient antibody

Increase the antibody concentration to all experimental tubes.

Intracellular target not available

Check to see if the target protein is intracellular. If so, make sure to permeabilize on ice and with cold reagents in order to stop all reactions.

Excess cell number

Dilute sample tubes in PBS for recommended concentration

Lasers not aligned

Run setup beads and adjust alignment if necessary.

Soluble/secreted target protein

Target protein must be membrane bound or cytoplasmic to be detected. A Golgi-block step, such as Brefaldin A, may improve signal.

Offset too high/gain too low

Make sure your positive control has a peak at 102. Check fluorescent signal and adjust voltage (within reason).

Primary and secondary are not compatible

Secondary should be raised against the host species of the primary.

PE antibody doesn't stain but FITC gives good results

Paraformaldehyde can release methanol in its breakdown, which will affect the staining.

Fluorochrome fluorescence faded

Sample may have been left out too long or been exposed too long. Experiment should be repeated with fresh antibody.

High fluorescence intensity

Antibody concentration

Too much antibody will give high non-specific binding or very high intensity fluorescence.

Excess antibody trapped

Ensure adequate washing steps and include tween or triton in the wash buffers. Fluorochrome molecules can be trapped while staining intracellular.

Inadequate blocking

Dilute antibodies in blocking solution, and increase the blocking time.

Two cell populations

More than one cell population present expressing target protein

Lessen the This may be in part to debris or adequate cell separation. Check expected expression levels from cell types contained in the sample.

Cell doublets present

Cells can be filter or sieved to remove unnecessary clumps. Mix cells gently with a pipette before staining and again before acquiring.
 

High side scatter background

Cells lysed

Samples need to be fresh and prepared correctly. Do not vortex too violently, or centrifuge too fast.

Bacterial contamination

Bacterial will auto fluoresce at low level. This will give a high event rate

Low event rate

Low number of cells/ml

Optimal cell concentration for each sample should be around 1x10^6 cells/ml. Ensure cells are gently mixed.

Cells clumped/blocking tubing

Pipette gently before staining and again before running. It's recommended to use to a filtered tube when running sample.

High event rate

Protein load

Use less protein and/or try an immunoprecipitation to enrich your target in the lysate.

Antibody concentration

Dilute the antibody concentration.

Unusual scatter profiles

Showing dead cells

This may be debris, and gates need to be adjusted. Or all cells could have died during mixing or additional chemicals added. Repeat experiment with fresh cells.

Activation methods may affect scatter characteristics of cells

This should be ok if the positive and negative control tubes were properly set up. Or if required, consult a different activation method.

Unexpected staining

Presence of dead cells

Always sieve the cells once before acquiring and sorting to remove any dead cell debris. Use freshly isolated cells as opposed to frozen cells whenever possible.

Affect antigen detection

Permeablization is required. However, reagents can affect certain antigens, i.e. lysis buffers and EDTA affects platelet markers.

Loss of epitope

Too much paraformaldehyde

Make sure to only use 1% paraformaldehyde.

Sample fixed too long

Follow proper instructions, most cells only need to be fix for less than 15 minutes. It is recommended to not fix for too long, as this will damage cells.

Sample wasn't kept cold

Follow proper protocol instructions, however most antibodies need to be kept at 4C to prevent loss of activity. Also, this prevents active phosphatases and proteases from altering the epitope of interest.

High background/high percentage of cells

Gain set too high/offset too low

Use the positive control to set up the flow cytometer correctly again, using the offset to reduce background from small particles and reduce the gain to decrease the signal.

Excess antibody

Decrease the antibody concentration. Detergent can also be added to the wash buffers to ensure any excess antibody is washed away.

No events

Clogged sample tube

Unclog flow cytometer injection tube as per instrument manufacturer's instructions (typically run 10% bleach for 5-10 min, followed by dH2O for 5-10 min).